Skip to content

Medical Devices · Process Validation

Medical Device Process Validation Consulting

The short answer

Device manufacturers must validate any process whose results cannot be fully verified by subsequent inspection and test. The judgment call — what genuinely requires validation versus verification — is where most programs go wrong, and it is where we usually start.

01

Why this matters

  • Over-validation consumes capacity; under-validation puts product conformity on inspection alone, which is rarely defensible for processes like welding, molding or sterilization.
  • OQ challenge conditions that never leave nominal settings do not establish a process window.
  • Software used in production or the quality system requires validation for its intended use, and this is a frequently cited gap.
02

Applicable regulations and standards

21 CFR 820.75 / QMSR
Process validation where results cannot be fully verified by subsequent inspection.
ISO 13485 clause 7.5.6
Validation of processes for production and service provision.
GHTF/SG3 process validation guidance
Decision framework and IQ/OQ/PQ methodology widely used by industry and auditors.
ISO 11135 / 11137
Sterilization validation for ethylene oxide and radiation processes.
03

What our consultants actually do

  • Build the validation decision tree that documents which processes are validated, verified or both
  • Write and review IQ, OQ and PQ protocols with challenge conditions that establish an actual process window
  • Support sterilization, packaging and seal validation, including aging and distribution simulation
  • Validate production and QMS software for intended use, with risk-based test depth
  • Define revalidation triggers and connect them to change control
  • Remediate legacy validation packages, including inherited processes from acquisitions or contract manufacturers
04

Across the product lifecycle

  1. 01 / DEVELOP

    Process capability considered during design, not after design freeze.

  2. 02 / SCALE

    Tooling, fixtures and automation qualified before production commitment.

  3. 03 / TRANSFER

    Design transfer evidence that the process can be run in production.

  4. 04 / VALIDATE

    IQ/OQ/PQ executed with justified sample sizes and challenge conditions.

  5. 05 / MANUFACTURE

    Monitoring, control charts and nonconforming product handling.

  6. 06 / MAINTAIN

    Revalidation on change, periodic review and equipment lifecycle.

05

Problems we are usually called about

  • Sample sizes chosen by convention with no confidence and reliability rationale
  • OQ run only at nominal parameters, so no proven acceptable range exists
  • Contract manufacturer validation accepted without review of the underlying data
  • Software validation limited to installation testing
  • Change control that does not evaluate validation impact
06

Typical deliverables

  • Validation master plan and process validation decision matrix
  • IQ/OQ/PQ protocols with statistical rationale
  • Sterilization and packaging validation strategy
  • Software validation plans and test records
  • Legacy validation gap assessment and remediation sequence
07

Frequently asked

Which device processes must be validated?
Any process whose output cannot be fully verified by subsequent inspection and test — commonly sterilization, sealing, welding, molding, coating, cleaning and many automated assembly steps. The decision should be documented, including for processes you conclude do not require validation.
How are sample sizes justified?
Through a stated confidence and reliability basis tied to the risk of the attribute being evaluated, documented in the protocol before execution.
Do we need to revalidate after moving equipment?
A relocation should trigger an assessment through change control. In most cases at least installation qualification is repeated, with operational or performance qualification scoped by risk.
08

References

Start a project

Tell us what you're working on.

Describe the product, the process or the problem. We will tell you honestly whether we are the right people to help, and who should be on the team if we are.

CallStart a Project